Aerospace Heat Treatment Thermocouples to AMS 2750D

Aerospace Heat Treatment applications require AMS 2750D thermocouples. The key point of this standard, for sensor makers, demands that Thermocouples are calibrated by approved means and shown to be within defined limits. Individual sensors or batches of Thermocouple material can be tested. The maximum acceptable error depends on sensor use. There are strict limits on the number of uses before recalibration or replacement is required.

 

Peak Sensors often supplies Mineral Insulated Thermocouples, Rare Metal Thermocouples or Thermocouple wire batch certified. Thermocouple type R, type S and type N are the most common aerospace sensors.

 

Cost Reduction for AMS 2750D

 

To minimise Mineral Insulated Thermocouple sensor costs, Peak Sensors will have batches of wire calibrated and then from it, build sensors according to a customers requirement.… Read more

Temperature Sensor Radiation Error in an Aerospace Furnace

Measurement errors can be undetected and highly significant. I had a great example of sensor radiation error recently.

I took a telephone call from a customer telling me my temperature sensors were inaccurate. He was building an AMS 2750D compliant Aerospace Furnace. He bought a number of thermocouples from Peak Sensors which were all UKAS calibrated.

System Design

There was a combination of Mineral Insulated Type N and Ceramic Insulated Type R (wire Ø 0.25mm) sensors. In four locations, a type N and a type R were installed adjacent to each other. One would expect them to read very similar temperatures. During pre commissioning trials it was noticed that at times there was up to 50°C difference between the pair.… Read more